Приказ основних података о документу

MSV signal processing system for neutron-gamma discrimination in a mixed field

dc.creatorStanković, Srboljub J.
dc.creatorIričanin, Bratislav D.
dc.creatorNikolić, Dragana
dc.creatorJanković, Ksenija
dc.creatorRadenković, Mirjana
dc.creatorStanković, Koviljka Đ.
dc.creatorOsmokrović, Predrag V.
dc.date.accessioned2022-04-18T15:08:27Z
dc.date.available2022-04-18T15:08:27Z
dc.date.issued2012
dc.identifier.issn1451-3994
dc.identifier.urihttp://rims.institutims.rs/handle/123456789/191
dc.description.abstractNa osnovu principa koji su proizišli iz Kempbelove teoreme, u ovom radu je urađena analiza nekih mogućnosti Kempbelovog MSV sistema za obradu signala. MSV mod je posebno pogodan za merenja koja se sprovode u mešovitom polju zračenja, kada su značajno različite količine naelektrisanja u detektoru koja potiču iz interakcija dva tipa zračenja. Merni detektorski element za mešovita n-γ polja mogao bi da bude odgovarajuća jonizaciona komora ili poluprovodnička komponenta. Sprovedeno je ispitivanje diskriminacije gama komponente u odnosu na neutronsku komponentu izlaznog signala iz detektora, pri čemu su urađeni proračuni u skladu sa teorijskim modelom interakcije zračenja sa detektorom. Potvrđena je prednost MSV metode i zaključeno je da je stepen n-γ diskriminacije prilikom MSV obrade signala veća nego kod klasične metode merenja.sr
dc.description.abstractBased on the principles derived from Campbell's theorem, this paper carries out an analysis of the possibilities of Campbell's mean square value signal processing system. The mean square value mode is especially suitable for measurements performed in a mixed radiation field, because the quantities of electrical charge involved in the interactions of the two types of radiation are substantially different. The measuring detector element may be an adequate ionization chamber and/or semiconductor components for mixed n-γ fields. An examination of the discrimination of gamma in relation to the neutron component in the signal of the detector output was carried out, calculated according to the theoretical model of radiation interaction with the detector. The advantage of the mean square value method was confirmed and it was concluded that the order of n-γ discrimination in mean square value signal processing is greater than the one rendered by the classical measuring method.en
dc.publisherUniverzitet u Beogradu - Institut za nuklearne nauke Vinča, Beograd
dc.relationinfo:eu-repo/grantAgreement/MESTD/Basic Research (BR or ON)/171007/RS//
dc.relationinfo:eu-repo/grantAgreement/MESTD/Technological Development (TD or TR)/37021/RS//
dc.relationinfo:eu-repo/grantAgreement/MESTD/Integrated and Interdisciplinary Research (IIR or III)/43009/RS//
dc.relationinfo:eu-repo/grantAgreement/MESTD/Integrated and Interdisciplinary Research (IIR or III)/41025/RS//
dc.rightsopenAccess
dc.rights.urihttps://creativecommons.org/licenses/by-nc-nd/4.0/
dc.sourceNuclear Technology & Radiation Protection
dc.subjectpoluprovodnički detektorsr
dc.subjectneutron-gama diskriminacijasr
dc.subjectjonizaciona komorasr
dc.subjectsemiconductor detectoren
dc.subjectneutron-gamma discriminationen
dc.subjectionization chamberen
dc.titleMSV sistem za obradu signala sa neutron-gama diskriminacijom u mešovitom poljusr
dc.titleMSV signal processing system for neutron-gamma discrimination in a mixed fielden
dc.typearticle
dc.rights.licenseBY-NC-ND
dc.citation.epage170
dc.citation.issue2
dc.citation.other27(2): 165-170
dc.citation.rankM22
dc.citation.spage165
dc.citation.volume27
dc.identifier.doi10.2298/NTRP1202165S
dc.identifier.fulltexthttp://rims.institutims.rs/bitstream/id/69/188.pdf
dc.identifier.scopus2-s2.0-84864237783
dc.identifier.wos000306036600011
dc.type.versionpublishedVersion


Документи

Thumbnail

Овај документ се појављује у следећим колекцијама

Приказ основних података о документу